Randomized Trial of a Distal Embolic Protection Device During Percutaneous Intervention of Saphenous Vein Aorto-Coronary Bypass

Donald S. Baim, MD, Dennis Wahr, MD, Barry George, MD, Martin B. Leon, MD, Joel Greenberg, MD, Donald E. Cutlip, MD, Unsal Kaya, MS, Jeffrey J. Popma, MD, Kalon K.L. Ho, ,MD, MSc, Richard E. Kuntz, MD, MSc; on behalf of the Saphenous vein graft Angioplasty Free of Emboli Randomized (SAFER) Trial Investigators
August 2002
ABSTRACT: Background. Stents provide effective treatment for stenotic saphenous venous aorto-coronary bypass grafts, but their placement carries a 20% incidence of procedure-related complications, which potentially are related to the distal embolization of atherosclerotic debris. We report the first multicenter randomized trial to evaluate use of a distal embolic protection device during stenting of such lesions. Methods and Results. Of 801 eligible patients, there were 406 randomly assigned to stent placement over the shaft of the distal protection device, and 395 were assigned to stent placement over a conventional 0.014´´ angioplasty guidewire (control group). The primary endpoint — a composite of death, myocardial infarction, emergency bypass, or target lesion revascularization by 30 days — was observed in 65 patients (16.5%) assigned to the control group, and 39 patients (9.6%) assigned to the embolic protection device (p = 0.004). This 42% relative reduction in major adverse cardiac events was driven by myocardial infarction (8.6% versus 14.7%; p = 0.008) and “no-reflow” phenomenon (3% versus 9%; p = 0.02). Clinical benefit was seen even when platelet glycoprotein IIb/IIIa receptor blockers were administered (61% of patients), with composite endpoints occurring in 10.7% of protection device patients versus 19.4% of control patients (p = 0.008). Conclusions. Use of this distal protection device during stenting of stenotic venous grafts was associated with a highly significant reduction in major adverse events compared with stenting over a conventional angioplasty guidewire. This demonstrates the importance of distal embolization in causing major adverse cardiac events and the value of embolic protection devices in preventing such complications. Catheter-based intervention in saphenous venous aortocoronary bypass grafts carries a significant (? 20%) risk of a major adverse clinical event (MACE) (predominantly myocardial infarction) or reduced antegrade flow (the noreflow phenomenon).1 Several mechanisms have been offered, including spasm of the distal microcirculation, platelet clumping, and most recently, the distal embolization of pieces of friable lipid-rich plaque.2 Preliminary work with the PercuSurge GuardWire — a device for transient distal balloon occlusion during angioplasty or stent placement that allows recovery of any liberated plaque by aspiration before restoration of antegrade flow — has demonstrated consistent recovery of plaque constituents (cholesterol crystals, foam cells, fibrous plaque) that otherwise would have embolized into the myocardial bed.3 This initial experience has also suggested a reduced incidence of myocardial infarction (4 The Saphenous vein graft Angioplasty Free of Emboli Randomized (SAFER) trial was an 801-patient U.S. multicenter study, in which patients undergoing saphenous vein graft intervention were randomized to undergo either stenting with a conventional guidewire or stenting with the GuardWire distal protection device. The SAFER trial was the pivotal trial that led to U.S. Food and Drug Administration approval in August 2001. Methods The primary objective of Saphenous vein graft Angioplasty Free of Emboli Randomized (SAFER) trial was to compare the 30-day clinical outcome after saphenous vein graft stenting plus GuardWire distal protection versus that performed over a conventional guidewire (control arm). This randomized trial complied with the Declaration of Helsinki with regard to investigation in humans, was approved for Investigational Device Exemption by the U.S. Food and Drug Administration, and was approved by the local hospital Institutional Review Boards at each of the investigational sites. Eligibility criteria. Patient candidates had a history of angina and signs of myocardial ischemia resulting from a target lesion > 50% diameter stenosis (angiographic visual assessment) located in the mid-portion of a saphenous vein graft, with a reference diameter between 3 and 6 mm. In the first 142 patients, the lesion could not occupy more than one-third of the graft length. In subsequent patients, no upper limit on lesion length was imposed. Major exclusion criteria included: 1) recent myocardial infarction with baseline elevation of cardiac enzymes (creatine kinase-MB fraction); 2) significantly impaired left ventricular function (ejection fraction 2.5 mg/dL (unless on long-term hemodialysis); and 4) planned use of an atherectomy device. Coronary intervention. After informed consent, patients were premedicated with aspirin (325 mg orally) and brought to the interventional laboratory. During the procedure, intravenous heparin was administered to prolong the activated clotting time to > 250 seconds. A platelet glycoprotein IIb/IIIa receptor blocker was used at the discretion of the operator. Subjects were randomized to undergo stenting performed over either a conventional 0.014´´ angioplasty guidewire or a 0.014´´ PercuSurge GuardWire balloon occlusion device, with randomization stratified by site and by whether the operator preselected IIb/IIIa receptor blockade. The series of treatments for both arms of the study involved optional pre-stent dilatation of the lesion, deployment of >= 1 stent, and optional post-stent dilatation (at higher pressure or with a larger diameter balloon). In patients assigned to the GuardWire arm (Figure 1), the 0.014´´ hollow-core GuardWire was advanced across and beyond the target lesion and was attached to a proximal adaptor that allowed progressive inflation of the elastomeric balloon at its tip (range of inflated diameter, 3–6 mm) with dilute radiographic contrast until the antegrade flow of contrast within the graft was halted. The lumen of the GuardWire was then sealed, allowing removal of the adaptor and serial performance of the stent procedure (using the GuardWire shaft in lieu of a conventional guidewire). After satisfactory stent deployment, a 5 French (1.7 mm) diameter aspiration catheter (Export) was advanced over the GuardWire until it lay just proximal to the occlusion balloon and was connected to an evacuated 20 cc syringe. Between 20–40 mL of blood was vigorously aspirated through this catheter before the adaptor was reattached and the distal occlusion balloon was deflated to restore antegrade flow. After satisfactory stent deployment, final angiograms were obtained. Standard post-stent therapy (aspirin 325 mg/d, clopidogrel 300 mg oral load, followed by 75 mg/day for 2–4 weeks) was commenced. Serial 12-lead ECGs were performed after the procedure and then daily until discharge; blood samples for measurement of serum creatine kinase (CK) and its myocardial (MB) fraction were collected after the procedure and every 8 hours thereafter until discharge. Data collection and core laboratory analysis. Detailed case report forms were completed by the clinical coordinators at each site, monitored by independent study monitors, and submitted to the data-coordinating center (CDAC/Harvard Clinical Research Institute, Harvard Medical School, Boston, Massachusetts). Angiograms obtained during the procedure were submitted to the angiographic core laboratory (Brigham and Women’s Angiographic Core Laboratory, Boston, Massachusetts), where they were analyzed with a computer-based system (Medis; Leiden, the Netherlands). The diameter of the reference coronary and the minimum lumen diameter of the target lesion were determined before the procedure, immediately after the procedure, and at follow-up. Study endpoints and statistical methods. The primary endpoint of the study, MACE rate at 30 days, was defined as the composite of death, myocardial infarction, emergent bypass surgery, or target vessel revascularization within 30 days of the index procedure. Death was defined as the occurrence of death from any cause. Myocardial infarction was defined as the occurrence of an elevated CK-MB fraction > 3 times the upper limit of normal (standardized to each clinical site’s normal range) in at least 1 of 3 serial protocol-driven cardiac enzyme measurements performed during the first 18–24 hours after the index procedure or in any subsequent clinically-driven measurement. Patients with enzymatic elevation were further divided into those with and without the appearance of pathological Q waves on serial ECGs. A clinical events committee, blinded to treatment assignment, determined all clinical endpoints. Technical success for patients assigned to the GuardWire arm was defined as delivery of the GuardWire system to the intended target site, followed by successful inflation, aspiration, and deflation, according to the Instructions for Use. Other prespecified secondary endpoints included acute thrombosis, postprocedure flow, and vessel injury (distal dissection or perforation). The study was designed to reject the null hypothesis (i.e., that there was no difference between the treatment groups) with a 2-tailed 5% level of significance and 80% power. On the basis of data from prior single-center vein graft intervention study,4 it was assumed that the 30-day primary endpoint rate would be 16% in the control arm and Baseline demographics. The baseline demographics are shown in Table 1, reflecting the advanced age, severe angina, and multiple risk factors common in trials of saphenous aorto-coronary vein graft intervention. Angiographic findings.The baseline and postprocedural angiographic data are shown in Table 2. They are noteworthy for large graft diameter (mean, 3.4 mm), long lesion length (mean, 16 mm; range, up to 79 mm), and the common presence of lesion-associated thrombus (39%). Procedural details. Stenting was performed in 848 of 875 (96.9%) lesions, using either balloon-expandable or self-expanding (9.6% for GuardWire-assigned patients and 20.4% for control patients) designs. The mean number of stents per lesion was 1.35 and 1.38 in the GuardWire and control groups, respectively. Most involved “primary” stenting — that is, placement of the stent without predilatation (79.4% in the GuardWire group and 67.7% in the control group). Postdilatation after stent deployment was performed in 27.3% of GuardWire patients and 40% of the control patients, using a mean balloon size of 4.20.7 mm and a mean maximum inflation pressure of 13.44.5 atm. Residual minimum lumen diameter (3.220.63 mm in patients stented with the GuardWire, and 3.200.58 in control patients) and diameter stenosis (5.913.3% in patients stented with the GuardWire and 6.111.3% in control patients) were similar (Table 3). In the GuardWire arm, technical success was achieved in 90.1% of cases (henceforth designated as the per-protocol cohort). Technical failures included inability to deliver the GuardWire to the intended location (5.4%), inability to achieve or sustain occlusion of antegrade flow (3.2%), and inability to perform at least 20 mL of aspiration before deflation (1.2%). The median GuardWire occlusion time was 388 seconds (25th and 75th percentiles, 265 and 640 seconds). Endpoint results. There was a 6.9% absolute (42% relative) reduction in the 30-day primary endpoint (9.6% for GuardWire patients versus 16.5% for control patients; p 0.004) (Figure 2 and Table 4). This predominantly reflects a reduction in myocardial infarctions of all magnitudes (Figure 3). In addition, rates of TIMI grade 3 flow were higher for the GuardWire arm (98%) compared with the control arm (95%; p = 0.04; Table 3), and the incidence of clinically evident no-reflow was reduced (3% versus 9%; p = 0.001). There was no significant increase in subacute closure of the treated graft (1.7% for GuardWire versus 0.5% for conventional treatment; p = 0.18) and a lower rate of perforation for the GuardWire arm (0.2% versus 1.5%; p = 0.05) (Table 3). Beyond this intent-to-treat primary endpoint, a perprotocol analysis was performed on 90.1% of GuardWire patients with technically successful use of the device. This showed an even lower incidence of the primary endpoint of myocardial infarction (7.9%) (Figure 3) and no-reflow phenomenon (2.4%). The rates of primary endpoint (25.0%) and no-reflow (12.8%) in those with technical failure of the GuardWire arm were similar to the control arm. Patients who were preselected for use of a platelet GP IIb/IIIa receptor blocker [232 of 406 patients (57%) in the GuardWire arm, and 232 of 395 patients (58%) in the control arm] had a higher incidence of MACE than patients selected not to receive IIb/IIIa blockers (Table 4). This suggests operator selection for these agents in patients with higher risk lesion morphology. However, even when a GP IIb/IIIa receptor blocker was used, a clear reduction in MACE was seen with GuardWire distal protection (10.7% versus 19.4%; p = 0.008) (Table 4). Discussion On the basis of perfusion models of rabbit hind-limb and intraoperative human coronary arteries, the process of balloon angioplasty was known to involve radial vessel expansion and fracturing of intimal plaque, with only rare distal embolization of plaque constituents.6 Evidence for more routine distal embolization during angioplasty and stent placement, though, has now emerged from 2 sources: 1) recording of echogenic material by middle cerebral artery transcranial Doppler during carotid stent placement,7 and 2) recovery of atherosclerotic plaque debris when angioplasty is performed using one of the distal protection devices now in clinical testing.2,3 Those devices include various filters mounted near the end of conventional guidewires, as well as the occlusion/aspiration system used in the present study. It remains unclear whether these distal emboli are linked causally to adverse procedural events or whether their recovery would reduce the frequency of such events without producing other complications (i.e., thrombosis, distal vessel injury, etc.). The model of saphenous vein graft intervention is particularly apt for testing these issues. Between 7–10 years after bypass surgery, more than half of such grafts develop significant narrowing or occlusion.8 These narrowed grafts are commonly approached with catheter-based techniques (especially stenting) in an effort to avoid a repeat bypass surgery. Although high degrees of short-term success and low rates of in-stent restenosis have been achieved, the soft and friable nature of the lipid-rich plaque in such grafts contributes to the high occurrence of adverse clinical events (reduced flow despite a patent vessel, or periprocedural myocardial infarction), which are associated with increased 30-day and 1-year mortality in the this population.1,4 Given the frequent presence of platelet thrombi in such grafts, serotonin released by such platelets might cause distal microvascular (arteriolar) spasm,1 which is consistent with the observation that selective arteriolar vasodilators (calcium channel blockers, adenosine, nitroprusside) frequently improve or normalize episodes of reduced flow during vein graft intervention. A second proposed mechanism is that platelet aggregation itself might cause or amplify distal embolization. However, no consistent benefit has been seen with the use of potent antiplatelet agents during saphenous vein graft intervention.9 The focus on causation has shifted to distal embolization of atherosclerotic debris itself. Webb and colleagues3 described use of the GuardWire distal protection system in saphenous vein grafts, finding almost universal recovery of atheroembolic particles and a fewer adverse events (A total of 4% forMACE versus 17% for historical controls). The present study now completes this hypothesis by showing not only that embolic particles are recovered, but that their recovery is associated with a major reduction in adverse clinical events compared with placement of stents without distal protection. Some residual MACE are still seen in the GuardWire arm, suggesting ongoing technical challenges in obtaining complete distal protection. Importantly, there was no offsetting increase in the incidence of complications (distal dissection, perforation, or abrupt closure) because of use of the low-pressure elastomeric occlusion balloon in the distal graft. Although the SAFER trial was not powered to show a significant reduction in mortality, it did show a mortality trend (1.0% versus 2.3%; p = 0.17) that parallels the significant reduction in the primary endpoint. Finally, the addition of distal protection offered similar benefits against MACE, whether or not the operator had decided to pretreat with a platelet glycoprotein IIb/IIIa receptor blocker. The SAFER study makes clear the importance of distal atheroembolization and the benefit of devices that prevent it during catheter-based intervention in saphenous vein grafts and potentially in other territories (e.g., native coronary, carotid, renal arteries) where distal embolization causes significant end-organ damage. Acknowledgment. This clinical trial was supported under a U.S. Food and Drug Administration Investigational Device Exemption for PercuSurge Corporation, a division of Medtronic, Inc.
1. Piana RN, Paik GY, Moscucci M, et al. Incidence and treatment of “no-reflow” after percutaneous coronary intervention. Circulation 1994;89:2514–2518. 2. Topol EJ, Yadav JS. Recognition of the importance of embolization in atherosclerotic vascular disease. Circulation 2000;101:570–580. 3. Webb JG, Carere RG, Virmani R, et al. Retrieval and analysis of particulate debris following saphenous vein graft intervention. J Am Coll Cardiol 1999;34:461–467. 4. Hong MK, Mehran R, Dangas G, et al. Creatine kinase-MB enzyme elevation following successful saphenous vein graft intervention is associated with late mortality. Circulation 1999;100:2400. 5. Geller NL, Pocock SJ. Interim analysis in randomized clinical trials: ramifications and guidelines for practitioners. Biometrics 1987;43:213–223. 6. Sanborn TA, Faxon DP, Waugh D, et al. Transluminal angioplasty in experimental atherosclerosis — analysis for embolization using an in vivo perfusion system. Circulation 1982;66:917–922. 7. Jordan WD, Voellingeer DC, Doblar DD, et al. Microemboli detected by transcranial Doppler monitoring in patients during carotid angioplasty versus carotid endarterectomy. Cardiovasc Surg 1999;7:33–38. 8. DeFeyter PJ, van Suylen R, de Jaegere PPT, et al. Balloon angioplasty for the treatment of lesions in saphenous vein bypass grafts. J Am Coll Cardiol 1993;21:1539–1549. 9. Ellis SG, Linkoff AM, Miller D, et al. Reduction in complications of angioplasty with abciximab occurs largely independently of baseline lesion morphology. J Am Coll Cardiol 1998;32:1619–1623.

Advertisement

Advertisement

Advertisement